Doric Instruments 1000 Pro Series Digital Panel meters are suitable User Manual
Page 21
20
1000-ACV
AC Volts
Equipment Required:
1. Precision AC voltage source with a resolution to 0.1mV and
an accuracy of ±0.1%
2. Interconnecting copper wire from the AC source to the
indicator.
3. Trimmer adjusting tool.
STEP 1.
Disconnect power from the unit and using copper
connecting leads, connect the output of the AC Voltage source to
terminals 5 and 6 of the indicator according to the chart on page 5.
STEP 2.
Remove the front panel lens and reconnect power and
allow for a 5-minute warm-up.
STEP 3.
Input 0.0 mV from the AC voltage source and adjust the
ZERO pot for a display of 0.0±1 count. (Refer to the figure on page
15 for the location of the zero and span pots).
STEP 4.
Input 190.0 mV at the measured frequency from the AC
source and adjust the SPAN pot for a reading of 190.0±1 count.
STEP 5.
Repeat steps
3 and
4 until no further adjustment is
needed.