Doric 473/475

Doric Model 473 and 475 Digital Temperature
Calibrators use custom VLSI integrated circuits
in a patented circuit design to deliver
exceptional accuracy, high reliability, and long
battery life in a truly portable, handheld
instrument. Large, touch-type switches and
precision adjustment control let you access
multiple functions quickly, and because you
can both simulate thermocouples or RTDs and
measure corresponding transmitter output with
one instrument, calibration is faster.

Other features that increase your productivity
include direct readings in both °C and °F,
without additional modules or conversion
tables. Independent, simultaneous zero and
span adjustment functions eliminate repetitive
readjustments of instruments being calibrated.
Rugged package design and complete fuse
protection insure reliable use, year in and
year out.

General Specifications

473 Series

475 Series

Inputs:

Thermocouple: K, J, T, mV DC, mA DC,

RTD: Pt. .00385, Pt. .00392, Cu 10

,

, mV DC, mA DC

Calibration Output:

Thermocouple: K, J, T, mV DC, mA DC,

RTD: Pt. .00385, Pt. .00392, Cu 10

,

, mV DC, mA DC

Input/Output Impedance:

Input Impedance:

Calibration Mode:

T/C and mV: 13M

Output impedance during set-up for calibration: 100M

between RTD

mA 11

or 50 K

(selectable)

terminals 1, 2, 3, and 4.

Output Impedance:

Output impedance: during loop operation with an external instrument or

Less than 10

up to 150nA

transmitter, ohmic value between terminals 1 and 2 and between
terminals 3 and 4 is equal to the pre-set ohmic value of the HI or LO
potentiometer, from 1

to 2K

. Terminals 1 and 4 and terminals 2

and 3 are connected in common.

Input Impedance: 10–20

with mA input; 180K

with mV input.

Indication Mode:
Source Impedance: 21.5K

, with a nominal excitation current flowing

from terminal 4 to terminal 3 of 172 µA for temperature ranges
(55µA for 0.1ohm range, 5.5µA for 1.0 ohm range).

Input Impedance: 15M

between terminals 1 and 2.

Temperature Stability:

Zero: ± 2 µV/°C, Span: ±0.2% rdg/°C

Zero: ± 2 µV/°C, Span: ±0.2% rdg/°C

Ref. Junction: ±1°C/°C

Stability with Time:

90 days add ±0.05% rdg.

90 days add ±0.05% rdg.

1 year add ±0.1% rdg.

1 year add ±0.1% rdg.

Noise Rejection:

CMV: 1000VAC or VDC, 1500V peak

CMV: 150VAC RMS; 200VDC

CMR: 140dB @ 50/60Hz, 0.1° range; 120dB @

CMR: 140dB @ 50/60Hz ±1Hz, 0.1° range; 120dB @ 50/60Hz ±1Hz,

50/60Hz, 1.0° range; 100ohms unbalance

1.0° range, 100ohms unbalance

NMR: 50dB @ 50/60Hz ±1Hz, 0.1° range; 35dB @

50/60 Hz±1Hz, 1.0° range

Isolation:

Can be used with isolated or non-isolated instruments

Can be used with isolated or non-isolated instruments. The RTD input/output
terminals are fully isolated from the mA /mV input terminals.

Weight:

1 lb. (454g)

1 lb. (454g)

Size (H x W x L)

2.75" x 3.75" x 8.3" (6.9 cm x 9.5 cm x 21.1 cm)

2.75" x 3.75" x 8.3" (6.9 cm x 9.5 cm x 21.1 cm)

Key Features

Measures and Simulates RTDs, Thermocouples, Ohms, Milliamps, and Millivolts

Calibrates Temperature Controllers, Indicators, Recorders, Transmitters

Long Battery Life

Lightweight, Handheld Versatility (carrying case included)

Hi/Lo Memory

Accuracy ±0.1%

94348.Doric473-5.1 3/19/99 2:11 PM Page 1