19

CALIBRATION PATH

The calibration is performed in Kg or Lb, according to the units of measure used upon entering
into the calibration path.

Operator Action

Display

1. Press ON/OFF button and immediately press and

hold HOLD/RELEASE button for 3-4 seconds

Calibration

Process

2. Press

ENTER

Enter load weight

600.0

3. Using

cd keys and/or the keypad, set the

calibration load to at least 200 lbs (for best results
and greater accuracy, use a 600 lb weight). Press
ENTER

Zero calibration
Please clear the scale

4. Please clear the scale platform and press ENTER

Zero calibration

Please wait

5. Please wait without touching or shaking the scale

until the zero calibration process is finished

Weight calibration

Put: XXX.X

6. Load the scale platform with the required weight

and press ENTER

Weight calibration

Please wait

7. Please wait without touching or shaking the scale

until the calibration process is finished

Calibration
factor: X.XXXXX

8. Press

ENTER

Calibration
Please clear the scale

9. Please remove weight from the scale platform

Rebooting
Please wait

10. Please wait until the scale resumes normal

operation

Health O Meter
Pro Plus