2-12 AUTO CALIBRATION

(1) When the displays are zero, press . key then lb/kg key at the same time. Don‘t release your

fingers from the keys until the weight display is appeared “ - - - - - ”.

®

0

¬

3

.

.

Stable

3

the value of full capacity of the scale

(2) Put the mass (full capacity) on the platter. The scale will be calibrated automatically when it is

stable. While the scale is reverting to zero, please take the mass away from the platter.

¨

Press CE can escape from the calibration mode.

F

When the display appears the value of full capacity, you can directly input the value of the

weight which you want to calibrate. Then, place the mass (calibrated weight) on the platter.

The scale will be calibrated automatically when it is stable.

RECALL THE ORGINAL VALUES OF CALIBRATION

If there is something wrong with the auto calibration, press

CE

then lb/kg

at the same time.

Release your finger from the lb/kg then CE key while the scale is counting down. The scale will

return to the original values of which had been calibrated in factory.

¨

If scale with one weighing unit, please use key to instead of lb/kg key.

U.W.

PST