Safety Information

x

The 37XR-A Digital Multimeter is UL, CSA, and EN61010-1 certified for

Installation Category III – 600V and Category II – 1000V. It is recommended for
use with local level power distribution, appliances, portable equipment, etc,
where only smaller transient overvoltages may occur, and not for primary
supply lines, overhead lines and cable systems.

x

Do not exceed the maximum overload limits per function (see specifications)

nor the limits marked on the instrument itself. Never apply more than 1000V
dc/750 V ac rms between the test lead and earth ground.

x

Inspect the DMM, test leads and accessories before every use. Do not use any

damaged part.

x

Never ground yourself when taking measurements. Do not touch exposed

circuit elements or test probe tips.

x

Do not operate the instrument in an explosive atmosphere.

x

Exercise extreme caution when: measuring voltage >20V // current >10mA // AC

power line with inductive loads // AC power line during electrical storms //
current, when the fuse blows in a circuit with open circuit voltage >1000 V //
servicing CRT equipment.

x

Always measure current in series with the load – NEVER ACROSS a voltage

source. Check fuse first. Never replace a fuse with one of a different rating.

x

Do not change the position of the Function/Range Switch while the MIN MAX,

feature is enabled. Erroneous readings will result.

x

Remove test leads before opening the Battery Cover or case.