Amprobe ACDC-100 Clamp-On-Multimeters User Manual
Page 12

11
Electrical (23 °C ± 5 °C) < 75% Rh
DC Voltage
RANgE
Accuracy
400.0 mV
± (0.3% + 3dgts)
4.000V, 40.00V, 400.0V
± (0.5% + 3dgts)
600V
± (1.0% + 4dgts)
Resolution : 0.1 mV on 400 mV range
nMRR : >50dB @ 50/60hz
CMRR : >20dB @ DC, 50/60hz, Rs=1kΩ
input impedance : 10MΩ, 30pf nominal (1000MΩ for 400.0mV range)
Max input : 600VDC/VAC rms
AC Voltage
RANgE
frequency
Accuracy
400.0mV
1)
50hz to 500hz
± (4.0% + 4dgts)
4.000V, 40.00V, 400.0V
50hz to 60hz
± (1.0% + 4dgts)
4.000V, 40.00V, 400.0V
60hz to 500hz
± (1.5% + 4dgts)
600V
50hz to 500hz
± (2.0% + 4dgts)
Resolution: 0.1 mV on 400 mV range
CMRR : >60db @ DC to 60hz, Rs= 1kΩ
input impedance: 10MΩ, 30pf nominal
Crest factor (ACDC-100 TRMS): < 1.6: at full scale; < 3.2: at half scale
Max input: 600VDC/VAC rms
1)
Selection by Range button manually, and is specified from AC 40mV (AC 60mV for ACDC-
100 TRMS) and up
Resistance
RANgE
Accuracy
400.0Ω
± (0.8% + 6dgts)
4.000kΩ, 40.00kΩ, 400.0kΩ
± (0.6% + 4dgts)
4.000MΩ
± (1.0% + 4dgts)
40.00MΩ
± (2.0% + 4dgts)
Resolution: 0.1 Ω on 400 Ω range
open Circuit Voltage : 0.4VDC typical
Continuity Tester
open Circuit Voltage:
0.4Vdc typical
Range:
400.0Ω
Accuracy:
± (1.5% + 6dgts)
Audible threshold:
between 10Ω and 120Ω.
Diode Tester
open Circuit Voltage
< 1.6 Vdc
Test Current (Typical)
0.4mA