Amprobe ACD-50NAV ACD-51NAV ACDC-52NAV ACD-53NAV ACD-54NAV Navigator-Clamps User Manual
Page 32

26
(9) Resistance & Continuity & Diode:
ACD-50NAV, ACD-51NAV, ACDC-52NAV, ACD-53NAV, ACDC-54NAV
Function
Range
Accuracy
Resistance
999.9 e
± (1.0% + 5dgt)
9.999 ke
± (1.0% + 3dgt)
99.99 ke
Continuity
999.9 e
± (1.0% + 5dgt)
Diode
0.40 ~ 0.80V
± (1.5% + 5dgt)
Overload protection : 1000Vrms
Max. Test Current : Approx. 0.5mA.
Maximum Open Circuit Voltage for e,
R
Approx. 3V
Maximum Open Circuit Voltage for diode : Approx. ±1.8V
Continuity check :
Internal sound activates if the resistance of the circuit under
test is less than 30e (approximately).
Continuity Indicator : 2 KHz Tone Buzzer
(10) Capacitance:
ACD-50NAV, ACD-51NAV, ACDC-52NAV, ACD-53NAV, ACDC-54NAV
Function
Range
Accuracy
Capacitance
3.999 μF
± (1.9% + 8dgt)
39.99 μF
399.9 μF
3999 μF
Overload protection : 1000Vrms
(11) Temperature :
ACD-51NAV, ACDC-54NAV
Function
Range
Accuracy
°C
-50 °C ~ 99.9 °C
± (1% + 2°C)
100 °C ~ 399.9 °C
400 °C ~ 1000 °C
± (1% + 1°C)
°F
-58 °F ~ 211.9 °F
± (1% + 4°F)
212.0 °F ~ 751.9 °F
752 °F ~ 1832 °F
± (1% + 2°F)
Overload protection : 1000 Vrms
• The above specification is assumed at the ambient temperature
stability within ±1°C. In addition, the temperature probe has
to be connected to meter for more than 1 hour before taking
measurement.
• The meter needs 2 hours for stability for ambient temperature
variation more than ±5°C.