Amprobe AC68C Clamp-Multimeter User Manual
Page 12

8
4k Hz
1 Hz ± (0.5 %rdg + 5 dgt) 1 V / 5 A
660 Vrms
/800
A
20 kHz* 0.01 Hz ± (0.5 %rdg + 5 dgt) 5 V
660 Vrms
* - Voltage input only
Resistance
Range Resolution Accuracy
Open
Overload
circ.volts
Prot.
4 k
Ω 1
Ω± (1 %rdg + 5 dgt) ≤ 0.5 VDC
660 Vrms
Diode Test
Range Resolution
Accuracy Open
Overload
circ.volts
Prot.
G
0.001 V ± (1 %rdg + 2 dgt) 3.0 VDC max 660 Vrms
Max short circuit current: 0.8 mA (typical)
Audible Continuity Test
Range
Accuracy Open circ.volts
Overload Prot.
40
Ω
≤ 0.5 V
3.0 VDC max
660 Vrms
Max short circuit current: 0.8 mA (typical)
Peak Hold Measurement, AC Voltage (50 Hz – 60 Hz)
Range Resolution Accuracy
Input
Overload
Imped.
Prot.
400 V
0.1 V ± (1.5 %rdg +10 dgt) 1 M
Ω 660
Vrms
600 V
1 V
± (1.5 %rdg +10 dgt) 1 M
Ω 660
Vrms
Peak Hold Measurement, AC Current (50 Hz – 60 Hz)
Range Resolution
Accuracy
Overload
Prot.
400 A
0.1 A
± (2 %rdg +10 dgt)
800 A
600 A
1 A
± (2 %rdg +10 dgt)
800 A
Troubleshooting
In case of malfunction during the operation of the meter, the
following steps should be performed in order to isolate the cause
of the problem:
1.
Check the battery.
2.
Review the operating instructions for possible mistakes in
operating procedure.
3.
Check the clamp against a known current source. Use the
voltage function to measure a known voltage (AC and DC).
4.
Check the test leads for continuity (voltage and
resistance). Except for the replacement of the battery,
repair of the clamp should be performed only by a Factory
Authorized Service Center or by other qualified instrument
service personnel. Front panel and case can be cleaned
with a mild solution of detergent and water. Apply
sparingly with a soft cloth and let dry completely before
using. Do not use aromatic hydrocarbons or chlorinated
solvents for cleaning.