Amprobe AC68C Clamp-Multimeter User Manual

Page 11

background image

7

P >

Safety: Meets EN61010-1 Cat III 600V. EN61010-2-03, UL3111-

1, CSA22.2, No 1010-1

EMC: Meets EN55011, EN61000-4-2,4
EMC: This product complies with requirements of the following
European Community Directives: 89/336/EEC (Electromagnetic
Compatibility) and 73/23/EEC (Low Voltage) as amended by
93/68/EEC (CE Marking). However, electrical noise or intense
electromagnetic fields in the vicinity of the equipment may
disturb the measurement circuit. Measuring instruments will also
respond to unwanted signals that may be present within the
measurement circuit. Users should exercise care and take
appropriate precautions to avoid misleading results when making
measurements in the presence of electronic interference.

Electrical Specifications

Accuracy is ± (%reading + nbr digits) at 23 °C ± 5 °C, < 80 %
R.H. The current error is specified within the largest circle that
can be drawn inside the jaw.
DC Current
Range Resolution Accuracy

Overload Prot.

400 A

0.1 A

± (2 %rdg + 5 dgt) 800 A

600 A

1 A

± (2 %rdg + 5 dgt) 800 A

AC Current, Trms
Range Resolution Accuracy

Overload

Freq.

Prot.

Response

400 A

0.1 A

± (2 %rdg + 5 dgt) 800 A

40-450 Hz

600 A

1 A

± (2 %rdg + 5 dgt) 800 A

40-450 Hz

Conversion: Ac coupled Trms
Crest Factor (Non-Sinusoidal) (50 – 60 Hz)
1 – 3:

± 0.5 %

3 – 5:

± 3 % 5 – 7: ± 6 %

DC Voltage
Range Resolution Accuracy

Input.

Overload

Imped.

Prot.

600 V

1 V ± (0.75 %rdg + 2 dgt) 1 M

Ω 660

Vrms

AC Voltage Trms – Frequency response: 40-450 Hz
Range Resolution Accuracy

Input

Overload

Imped.

Prot.

400 V 0.1 V ± (1.2 %rdg + 10 dgt) 1 M

Ω 660

Vrms

600 V 1 V ± (1.2 %rdg + 10 dgt) 1 M

Ω 660

Vrms

Conversion: Ac coupled Trms
Crest Factor (Non-Sinusoidal) (50 – 60 Hz)
1 – 3:

± 0.5 %

3 – 5:

± 3 % 5 – 7: ± 6 %

Frequency (Hz) – Autoranging for current and voltage
Range Resolution Accuracy Sensitivity

Overload

Prot.