Amprobe ACF-3000DM-A Current Transducer User Manual
Page 8
8
Connect the A C F - 3 0 0 0 D M - A output cable to your measuring instrument using
an adapter.
Set up the A C F - 3 0 0 0 - D M - A range to either 300 or 3000A respective to the
setup on the measuring instrument
E x a m p l e :
Range (Amps)
Proper Setup on the DM-II
Proper Setup on the
A C F 3 0 0 0 - D M - A
< 3 0 0
A C F 3 0 0 0
3 0 0
3 0 0
< 3 0 0 0
A C F 3 0 0 0
3 0 0 0
3 0 0 0
O u t p u t :
300ARange: 1.0mVAC/Amp
3000A Range: 0.1mVAC/Amp
Return the ACF-3000 range switch to the OFF position when measurement is
completed to preserve battery life.
M A I N T E N A N C E
Preventive maintenance primarily consists of cleaning the transducers and
cables to prevent surface contamination. Use a mild detergent and water to
clean the transducers and cables. Remove the detergent with clear water,
then wipe dry with a clean cloth.
N o t e : The use of solvents as cleaners is not recommended unless
thoroughly tested and found harmless to all surfaces and parts.
Do not submerse current transducer into water or other fluids.