Danger, Warning – Precision Medical EasyPulse5 Oxygen Conserving Regulator User Manual

Page 11

background image

10

3

. Slowly open cylinder valve counterclockwise, until completely open.

4. Turn dial on the Conserver until indicating pointer is aligned with the

prescribed setting.

5. Breathe through your nose.

NOTE:When on a pulse setting, there is flow or a pulse only at the beginning of

each breath. If you do not feel the pulse at the beginning of each breath,

check the setting. If there is still no pulse, turn dial to an equivalent

“Continuous” setting.

• Because each patient’s breathing pattern is different and the environment

varies, it may be difficult to feel some low setting pulses.

6. To remove the Conserver from cylinder:

Completely close oxygen cylinder valve.

• Turn the Conserver dial to any “Continuous” position.

• Wait for oxygen to stop flowing from Conserver.

Remove the Conserver from cylinder valve.

DANGER

NEVER attempt to remove the Conserver from a cylinder unless the

cylinder valve is closed.

WARNING

• When cylinder pressure is 500 psi (3447 kPa, 34.47 bar) and below, it

is recommended to change to a full oxygen cylinder.

• NO OXYGEN is delivered in between settings.

• To avoid injury to patient:

• ALWAYS confirm prescribed setting before administering to patient

and monitor flow on a frequent basis.

WARNING

• Use only Precision Medical, Inc. Carry Bag designed for the

EasyPulse

5

to prevent an oxygen enriched environment.

• DO NOT place the unit under clothing while in use.

• When the Conserver is in use a small amount of oxygen is vented. Wearing

the unit under clothing may saturate fabrics with oxygen and cause them

to burn rapidly if exposed to sparks or flame. It may take several hours

for oxygen levels in fabrics to return to normal.

• ALWAYS insert cylinder and the Conserver into bag, cylinder first with

gauge facing mesh. Reference Accessory photo, (pg. 13).