beautypg.com

Warning, Operating instructions – Precision Medical EasyPulse5 Oxygen Conserving Regulator User Manual

Page 10

background image

9

1. Position the cylinder so the oxygen cylinder valve outlet is pointing away

from the user and any other person (s).

2. Before installing the Conserver, “crack” cylinder valve by opening the valve

slightly to remove any foreign particles and then close valve.

NOTE: (For NO GAUGE (NG) Models ONLY) The Conserver should only

be used on cylinders with a contents gauge.

3. Be sure the Conserver is in the “OFF” position. Align “OFF” with the

indicating pointer .

4. Be sure the high pressure inlet connection has a seal.

DO NOT USE without a seal. (See Product Diagram)

NOTE: For optimum safety, use only a Precision Medical, Inc.

Inlet Seal (Viton O-ring). (See Parts List)

The Inlet Seal supplied with the Conserver is reusable.

DO NOT USE any other type of seal.

5. Attach the Conserver to the oxygen cylinder valve.

6. Hand tighten knob clockwise until tight.

OPERATING INSTRUCTIONS

1. Attach a

standard adult single lumen oxygen nasal cannula, no longer

than 7 feet (2.1 m) to the Conserver’s outlet according to the cannula

manufacturer’s instructions.

WARNING

• DO NOT use pediatric, low flow nasal cannulas or oxygen masks

with the Conserver.

• DO NOT block the cannula connection or kink cannula tubing when the

Conserver is in use, this may damage the Conserver.

2. Place the cannula over your ears and position the prongs in your nose as

instructed by your Provider or cannula manufacturer.

OR

This manual is related to the following products: