Installation warning, Caution – Precision Medical EasyPulse5 Oxygen Conserving Regulator User Manual
Page 9

8
INSTALLATION
WARNING
• Read this User Manual before installing or operating the
Oxygen Conserving Regulator.
• Inspect the Conserver and cylinder valve to ensure they are
free of oils, greases or other contaminants.
• DO NOT direct flow of oxygen at any person, or flammable
material when cracking open the cylinder.
• DO NOT use without yoke seal.
• DO NOT use plastic yoke seals.
• Use Precision Medical, Inc. yoke seal made of viton and brass.
• The yoke seal supplied with the Conserver is reusable.
CAUTION
• Inspect the Oxygen Conserving Regulator for visual damage
before use, DO NOT USE if damaged.
• DO NOT use tools to tighten T-handle, this can lead to over
tightening and will cause damage to the Conserver.
• Be sure all connections are tight and leak free.
• DO NOT use liquid leak detector to test for leaks.
NOTE: For operational instructions on how to operate the
cylinder consult your Provider.
1. Position the cylinder so the oxygen cylinder valve outlet is pointing
away from the user and any other person(s).
2. Before installing the Conserver, “crack” cylinder valve by opening the
valve slightly to remove any foreign particles and then close valve.
3. Be sure the Conserver is in the “OFF” position. Align “OFF” with the
indicating pointer .
4. Be sure the CGA 870 connection high pressure inlet has a yoke seal.
5. Align Index Pins with Index Pin Alignment Holes in Post Valve.
6. Hand tighten T-handle until snug.