Specifications – Precision Medical EasyPulse5 Oxygen Conserving Regulator User Manual
Page 7

6
SPECIFICATIONS
Inlet Pressure Range:
300 - 3000 psi (2068 - 20684 kPa)
Pressure Gauge Accuracy:
3 - 2 - 3% of full scale
Dimensions: (Are approximate and may vary by model)
Model: 198705
Weight:
9.9 oz (279 g)
Overall Length:
4.75 in (12.07 cm)
Width:
2.63 in (6.67 cm)
Height:
2.50 in (6.35 cm)
Pulse Settings:
1, 2, 3, 4, & 5 l/min Equivalents
Flow Settings:
2 l/min Continuous (198705)
1 through 6 l/min Continuous (198712)
Accuracy: Continuous Settings: within ± 10%
Cannula Requirement:
Maximum 7 foot (2.1 m) long standard
adult single lumen nasal cannula.
Operating Conditions:
Temperature:
35°F to 105°F (1.7°C to 40.6°C)
Altitude:
Sea level to 10,000 ft (0 to 3,048 m)
Storage Conditions:
Temperature:
-40°F to 140°F (-40°C to 60°C)
Maximum Humidity:
95% Noncondensing
Oxygen Cylinder Connection:
CGA 870 Valve
Ignition and Fault Tolerance:
Meets ASTM G175-03
Specifications are subject to change without prior notice. Manuals available on our Website; www.precisionmedical.com
P
U
L
S
E
O
N
L
Y
Accuracy: Pulse Settings: Within +/- 15% of the nominal bolus value (at
each breath rate)
Savings Ratio:
Up to 5.7:1
Trigger Method:
Inspiratory effort (Negative pressure
from patient inhalation)
Breathing Frequency:
Up to 35 b/min