Precision Medical Continuous Subglottic Suction Regulator User Manual
Page 5

4
PM9100 Continuous
Subglottic Suction Regulator
SPECIFICATIONS
continued
weight:
0.38 lbs
(1.7 kg)
Shipping weight:
0.53
lbs
(0.24 kg)
Setting Range for PM9100 Series: 20-70 mmHg
Vacuum Ports:
Vacuum Connection: 1/8” NPT
Patient Connection: 1/4” Hose Barb with Flow Indicator
Operation:
OFF - No Vacuum
REG. - (Regulated) provides an adjustable, continuous
vacuum level
PUSH TO CLEAR - Temporarily increases vacuum level to line
vacuum.
Flow:
Model
Mode
Max Flow
PM9100 Series
REG
20-85 mmHg
13 lpm @ 20 mmHg up to 85 mmHg
37 lpm @ mAx of 85 mmHg
Push to Clear
54 lpm
mAxImUm FLOw IS mEASURED @ A VACUUm SOURCE OF 21 inHg
Maximum Vacuum PM9100 Series: REG. mode 78-85 mmHg
Operating Environmental Limits:
0°F to 122°F (-18°C to 50°C)
Recommended Environmental Operating Limits:
55°F to 85°F (13°C to 29°C)
Storage Environmental Limits:
Temperature Range: -4°F to 140°F (-20°C to 60°C)
Humidity: max 95% Noncondensing
Specifications are subject to change without prior notice.