Disassembly instructions, Assembly instructions – Precision Medical Continuous Vacuum Regulator User Manual
Page 7

6
REPAIR KITS
Analog Part#
Digital Part#
Pm3000 / Pm3000D Vac Reg
Rk6000
Rk6000D
Pm3000E / Pm3000DE Vac Reg
Rk6000E
Rk6000DE
Pm3000HV / Pm3000DHV Vac Reg
Rk6000HV
Rk6000DHV
Pm3000DEHV Vac Reg
N/A
Rk6000DEHV
Pm3100 / Pm3100D Vac Reg
Rk6100
Rk6100D
Pm3100E / Pm3100DE Vac Reg
Rk6100E
Rk6100DE
Pm3100HV / Pm3100DHV Vac Reg
Rk6100HV
Rk6100DHV
Pm3100EHV / Pm3100DEHV Vac Reg
Rk6100EHV
Rk6100DEHV
Pm3200E Vac Reg
Rk6200E
N/A
Pm3500 / Pm3500D Vac Reg
Rk6500
Rk6500D
Pm3500E / Pm3500DE Vac Reg
Rk6500E
Rk6500DE
Pm3600 / Pm3600D Vac Reg
Rk6600
Rk6600D
Pm3600E / Pm3600DE Vac Reg
Rk6600E
Rk6600DE
Not all Repair Kits are listed above, contact Precision Medical, Inc. for
availability.
DISASSEMBLY INSTRUCTIONS
(reference “Parts descriPtion”)
1. Loosen the Set Screw (Item# 11) in Selector knob.
2. Pull the Control knob Assembly (Item# 12) away from case. (The
Regulator module (Item# 10) is threaded onto the Control knob Assembly.)
3. Remove the screws (Item# 2) from the back of the Regulator.
4. Separate the Case Assembly (Item# 9) by pulling it away from the
Housing Assembly (Item# 1).
5. Remove the Selector Assembly (Item# 6) by pulling it away from the
Housing Assembly (Item# 1).
6. Remove the Gauge Assembly (Item# 3).
ASSEMBLY INSTRUCTIONS
1. to assemble, perform the “disassembLy instructions” in
reverse order.
NOTE: • ensure the selector assembly is inserted into the control
knob assembly with the groove in the 12 o’clock position.
• ensure tabs and slots on various components are properly
aligned and engaged when reassembling.
2. Lubricate all O-rings and cavities with Vacuum grease (part# 1775)
supplied in the Vacuum Regulator Repair kit.
3. repeat steps 1 through 3 of “oPeratinG instructions”.
4. Prior to returning Vacuum Regulator to service verify accuracy of gauge.