Disassembly instructions, Assembly instructions – Precision Medical Continuous Vacuum Regulator User Manual

Page 7

background image

6

REPAIR KITS

Analog Part#

Digital Part#

Pm3000 / Pm3000D Vac Reg

Rk6000

Rk6000D

Pm3000E / Pm3000DE Vac Reg

Rk6000E

Rk6000DE

Pm3000HV / Pm3000DHV Vac Reg

Rk6000HV

Rk6000DHV

Pm3000DEHV Vac Reg

N/A

Rk6000DEHV

Pm3100 / Pm3100D Vac Reg

Rk6100

Rk6100D

Pm3100E / Pm3100DE Vac Reg

Rk6100E

Rk6100DE

Pm3100HV / Pm3100DHV Vac Reg

Rk6100HV

Rk6100DHV

Pm3100EHV / Pm3100DEHV Vac Reg

Rk6100EHV

Rk6100DEHV

Pm3200E Vac Reg

Rk6200E

N/A

Pm3500 / Pm3500D Vac Reg

Rk6500

Rk6500D

Pm3500E / Pm3500DE Vac Reg

Rk6500E

Rk6500DE

Pm3600 / Pm3600D Vac Reg

Rk6600

Rk6600D

Pm3600E / Pm3600DE Vac Reg

Rk6600E

Rk6600DE

Not all Repair Kits are listed above, contact Precision Medical, Inc. for

availability.

DISASSEMBLY INSTRUCTIONS

(reference “Parts descriPtion”)
1. Loosen the Set Screw (Item# 11) in Selector knob.
2. Pull the Control knob Assembly (Item# 12) away from case. (The

Regulator module (Item# 10) is threaded onto the Control knob Assembly.)

3. Remove the screws (Item# 2) from the back of the Regulator.
4. Separate the Case Assembly (Item# 9) by pulling it away from the

Housing Assembly (Item# 1).

5. Remove the Selector Assembly (Item# 6) by pulling it away from the

Housing Assembly (Item# 1).

6. Remove the Gauge Assembly (Item# 3).

ASSEMBLY INSTRUCTIONS

1. to assemble, perform the “disassembLy instructions” in

reverse order.

NOTE: • ensure the selector assembly is inserted into the control

knob assembly with the groove in the 12 o’clock position.

• ensure tabs and slots on various components are properly

aligned and engaged when reassembling.

2. Lubricate all O-rings and cavities with Vacuum grease (part# 1775)

supplied in the Vacuum Regulator Repair kit.

3. repeat steps 1 through 3 of “oPeratinG instructions”.

4. Prior to returning Vacuum Regulator to service verify accuracy of gauge.