Specifications – Precision Medical Continuous Vacuum Regulator User Manual
Page 4
3
SPECIFICATIONS
GAUGE RANGE: *
Counterclockwise direction
P
M3000:
0 - 200 mmHg - Full Vacuum
PM3100HV:
0 - 300 mmHg - Full Vacuum
*PM3000E:
0 - 200 mmHg (0 - 26 kPa)
*PM3200E: 0 - 100 kPa (750 mmHg) - Full Vacuum
PM3000HV:
0 - 300 mmHg - Full Vacuum
PM3500:
0 - 150 mmHg
PM3100:
0 - 200 mmHg - Full Vacuum
*PM3500E:
0 - 200 mmHg (0 - 26 kPa)
*PM3100E:
0 - 200 mmHg (0 - 26 kPa)
PM3600:
0 - 760 mmHg
*PM3100EHV:
0 - 300 mmHg (0 - 40kPa)
*PM3600E:
0 - 760 mmHg (0 - 100 kPa)
GAUGE ACCURACY:
Analog:
± 5% of mAx
Digital/Analog, Dual Gauge:
Digital Display:
± 1% of Full Scale
Analog Gauge:
± 5% of mAx within ref. Indicator
VACUUM PORTS:
1/8 NPT Female
MODES:
PM3100, PM3200E3 & PM3600 Series:
PM3000, PM3200E & PM3500 Series:
LINE - Provides maximum, continuous
vacuum from the vacuum source
OFF - No Vacuum
REG. - (Regulated) provides an adjustable,
continuous vacuum level
OFF - No Vacuum
REG. - (Regulated) provides an adjustable,
continuous vacuum level
FLOW:
Models
Mode
Max Flow
*PM3000, PM3100, PM3200 & PM3500:
REG
51 l/min
LINE
55 l/min
**PM3600 SERIES:
REG
71 l/min
LINE
82 l/min
*
mAxImUm FLOW
is obtained with a vacuum source of 21” hg (71.1 kPa)
** mAxImUm FLOW
IS OBTAINED WITH A VACUUm SOURCE OF
25” hg (84.6 kPa)
MAxIMUM VACUUM:
PM3000 & PM3100:
REG. mode @ Full Vac - 396 mmHg (53 kPa)
PM3200:
REG. mode @ Full Vac - 100 kPa (750 mmHg)
PM3500D: Restricted to 170 mmHg (23 kPa)
± 10 mmHg (1.3 kPa)
PM3500E: Restricted to 160 mmHg (21 kPa)
PM3600:
REG. mode @ Full Vac - 760 mmHg (101 kPa)
Operating Environmental Limits:
0°f to 122°f (-18°c to 50°c)
Recommended Environmental Operating Limits: 55°f to 85°f (13°c to 29°c)
Storage Environmental Limits:
temperature range:
-4°f to 140°f (-20°c to 60°c)
Humidity:
max 95% Noncondensing
Battery:
3 Volt Lithium, AA (Digital Vacuum Gauge Models ONLY)
Specifications are subject to change without prior notice.