TROUBLESHOOTING

continued

Problem

Probable Cause

Remedy

Aspirator will not

operate when

connected to Auto DC

Power Cord

1. On/Off Switch in “OFF”

position

2. Car not turned on.

3. Auto DC Power Cord not

connected properly

4. Defective auto power port

receptacle

5. Defective Auto DC Power

Cord

1. On/Off Switch to “ON”

position

2. Turn on car to activate it’s

DC outlet

3. Check connections at both

ends of cord

4. Connect to a known working

auto power port

5. Contact your Provider

Battery will not charge

1. Not connected to AC power

source

2. Defective battery

1. Connect to AC power source

2. Contact your Provider

Unable to achieve

desired vacuum setting

1. Filter not connected properly

2. Filter not installed correctly.

3. Suction Canister not

connected properly

4. Defective Suction Canister

5. Defective Aspirator

6. Suction Tubing Leaks

1. Confirm and tighten

connections

2. Confirm orientation of filter.

3. Confirm and tighten

connections

4. Replace Suction Canister

5. Contact your Provider

6. Replace Suction Tubing

Inlet Filter becomes

blocked by fluid

(Hydrophobic only)

1. Collection Canister overflow

mechanism failed

1. Replace Collection Canister

2. Replace Filter

Aspirator stops

running, power switch

“ON”, battery or AC

power applied

1. Device overheating

1. Power “OFF” device and

allow cooling.

2. Ensure device has adequate

ventilation. Ensure vents are

not blocked.

ACCESSORY / REPLACEMENT PARTS

Only use accessories designed for use with this Aspirator.

• Use tubing intended for suction only.

• Use Collection Canisters designed for use with this Aspirator only.

• Use Precision Medical: Inlet Filter, External AC & DC Power Cord &

Carry Bag Only

Description

Part #

*Inlet Filter (Hydrophobic)

506851

Inlet Filter (Bacterial)

507754

800cc Canister, (Case-10 ea.)

502519-10

800cc Inlet Adapter Kit

507521

1200cc Canister

507522

Patient Suction Tubing

1955

Carry Bag

507263

AC Power Cord

506961

Auto DC Power Cord

507213

*Hydrophobic Filter will provide higher level of protection, stopping

the egress of liquid into the device, if canister overfill protection should fail.