Accessory / replacement parts, Troubleshooting – Precision Medical PM66 EasyGoVac Aspirator User Manual
Page 15
14
TROUBLESHOOTING
continued
Problem
Probable Cause
Remedy
Aspirator will not
operate when
connected to Auto DC
Power Cord
1. On/Off Switch in “OFF”
position
2. Car not turned on.
3. Auto DC Power Cord not
connected properly
4. Defective auto power port
receptacle
5. Defective Auto DC Power
Cord
1. On/Off Switch to “ON”
position
2. Turn on car to activate it’s
DC outlet
3. Check connections at both
ends of cord
4. Connect to a known working
auto power port
5. Contact your Provider
Battery will not charge
1. Not connected to AC power
source
2. Defective battery
1. Connect to AC power source
2. Contact your Provider
Unable to achieve
desired vacuum setting
1. Filter not connected properly
2. Filter not installed correctly.
3. Suction Canister not
connected properly
4. Defective Suction Canister
5. Defective Aspirator
6. Suction Tubing Leaks
1. Confirm and tighten
connections
2. Confirm orientation of filter.
3. Confirm and tighten
connections
4. Replace Suction Canister
5. Contact your Provider
6. Replace Suction Tubing
Inlet Filter becomes
blocked by fluid
(Hydrophobic only)
1. Collection Canister overflow
mechanism failed
1. Replace Collection Canister
2. Replace Filter
Aspirator stops
running, power switch
“ON”, battery or AC
power applied
1. Device overheating
1. Power “OFF” device and
allow cooling.
2. Ensure device has adequate
ventilation. Ensure vents are
not blocked.
ACCESSORY / REPLACEMENT PARTS
Only use accessories designed for use with this Aspirator.
• Use tubing intended for suction only.
• Use Collection Canisters designed for use with this Aspirator only.
• Use Precision Medical: Inlet Filter, External AC & DC Power Cord &
Carry Bag Only
Description
Part #
*Inlet Filter (Hydrophobic)
506851
Inlet Filter (Bacterial)
507754
800cc Canister, (Case-10 ea.)
502519-10
800cc Inlet Adapter Kit
507521
1200cc Canister
507522
Patient Suction Tubing
1955
Carry Bag
507263
AC Power Cord
506961
Auto DC Power Cord
507213
*Hydrophobic Filter will provide higher level of protection, stopping
the egress of liquid into the device, if canister overfill protection should fail.