Maintenance – Precision Medical PM66 EasyGoVac Aspirator User Manual
Page 10
9
CLEANING / MAINTENANCE
WARNING
• Turn the Aspirator “OFF” ( ) and disconnect from any external power
source before cleaning.
• DO NOT clean the Aspirator with Alcohol, Acetone or other solvents.
CAUTION
• Tubing is single patient use only. DO NOT clean.
• Canister is single patient use only.
MAINTENANCE
Vacuum/Flow verification should be checked by the device provider as
necessary - Recommended annually or between patients.
Cleaning for Exterior Components:
1. Disconnect the Aspirator from power source and other connections before
cleaning.
2. Use a cloth or sponge dampened with mild soap and water.
3. Wipe dry with a clean cloth or paper towel.
4. Store the Aspirator in a clean area free from grease, oil, and other sources
of contamination.
1200cc canister (reusable) can be cleaned with a mild detergent or placed on
top rack of dishwasher.
For institutional cleaning:
Exterior: Hospital approved quaternary cleaner; i.e., Sporociden®
Reusable canister: Cydex® or equivalent