Grounding instructions, Danger, Warning – Precision Medical PM61 PowerVac Aspirator User Manual
Page 7: Operating instructions warning, Caution

ASPIRATOR
6
GROUNdING INsTRUCTIONs
The Power Vac Aspirator must be plugged into an electrical outlet that is properly
grounded. In the event of an electrical short circuit, grounding reduces the risk
of electrical shock. This product is equipped with a grounding circuit.
dANGeR
Improper use of the grounding plug can result in a risk of electrical
shock.
wARNING
Ground reliability can only be achieved when the power cord is
connected to an equivalent receptacle marked "hospital Grade".
OPeRATING INsTRUCTIONs
wARNING
Read this User manual before operating the Power Vac Aspirator.
CAUTION
Inspect the Power Vac Aspirator for visual damage, dO NOT USE if
damaged.
1. Place Aspirator near a properly grounded electrical receptacle on a table,
floor or other horizontal, smooth, hard surface.
Allow a minimum of 4 in (10 cm) clearance on all sides to ensure
adequate air flow for cooling of Aspirator.
2. Ensure the on/off switch and selector Knob are in the “oFF” position.
3. Ensure the Inline Hydrophobic Filter is properly connected to the vacuum
port on Suction Canister lid.
• the Inline Hydrophobic Filter prevents liquid from entering
the vacuum pump
.
CAUTION
dO NOT run the Power Vac Aspirator without the Inline hydrophobic
Filter. This will void the warranty.
4. Plug the Aspirator into a properly grounded receptacle.
5. Press the on/off switch to the “ON” position.
6. Block the patient connection on the Suction Canister and adjust the
Vacuum Control Knob to the desired vacuum.
To INCREASE vacuum - Turn knob CLOCKWISE
To DECREASE vacuum - Turn knob COUNTERCLOCKWISE
• Vacuum should be set as prescribed by your Physician or Healthcare
Professional.