ASPIRATOR

9

RePLACemeNT PARTs

Part #

description

1690

Inline hydrophobic Filter

1955

Suction Tubing, 1/4 Id x 6 ft, Clear Non-Conductive PVC

502391

Small Air Filter

502471

Large Air Filter

502687

Suction Canister - 2,000 cc (with lid)

502690

Inline hydrophobic Filter & Tube Assy

503240

Fuse

TROUBLeshOOTING

If the Power Vac Aspirator fails to function, consult the Troubleshooting
Guide below. If problem cannot be solved, consult your Provider.

Problem

Probable Cause

Remedy

Aspirator will not

operate

1. On/Off switch in

“oFF” position

2. Power Cord not

connected properly

3. thermal limit

exceeded

4. Blown Fuse

1. Push On/Off switch to

“on” position

2. Check plug at outlet

3. shut Aspirator “oFF”,

wait 15 minutes, be sure

vents are clear

4. Replace Fuse

No vacuum

on gauge / no

vacuum at Patient

Connection

Tubing

1. hydrophobic Filter

assembly not

connected properly

2. Suction Canister

not connected or

assembled properly

3. defective suction

Canister

4. defective regulator

1. Confirm and tighten

connections

2. Confirm and tighten

connections/assembly

3. replace suction

Canister

4. Contact your Provider

No Vacuum on

gauge/vacuum

at Patient Tubing

Connection

1. defective vacuum

gauge

1. Contact your Provider

Vacuum on gauge

/ No vacuum at

Patient Tubing

1. Clogged hydrophobic

Filter

2. Canister overfilled

1. Replace hydrophobic

Filter

2. Empty canister

Unable to adjust

vacuum

1. defective regulator

1. Contact your Provider