Instruction Manual

245364-V
May 2002

3-6 Operation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 755A

ITEM

CONTROL

DESCRIPTION

1

Recorder Oxygen Span

Selection Jumper

Provides selectable span of 100%, 20%, 10%, 5%, 2%, or 1% oxygen for analog output

to recorder and alarms. Note that on the circuit board the jumper positions are marked

according to the amplifier gain.

2

Recorder Zero

Suppression Selection

Jumper

Used in combination with item 3 to establish required zero suppression to obtain desired

range for analog output to recorder and alarms. Jumper provides selectable zero

suppression of 20%, 40%, 60%, or 80%

3

Recorder Zero

Suppression Coarse

Adjustment R41

Continuously adjustable from 0% to 25% oxygen. Thus the total suppression range is

0% to 105% oxygen.

4

Recorder Zero

Suppression Fine

Adjustment R104

Adjustable range of 2%.

5

Recorder Voltage

Output Selection Jumper Provides selectable output of 10mV, 100mV, 1V or 5V for a voltage recorder.

6

Recorder Span

Adjustment R88

Provides

±5% span adjustment of recorder output.

7

CAL 1 Potentiometer

R98

Used in calibration of automatic pressure compensation.

8

CAL 2 Potentiometer

R99

Used in calibration of automatic pressure compensation.

9

Detector Coarse Zero

Adjustment R9

Provides coarse adjustment of detector zero by shifting the null position of the detector

within the magnetic field. It is adjusted at the factory and does not require readjustment

except after replacement of detector.

10

Response Time

Adjustment R30

Provides adjustment range of 5 to 25 seconds for electronic response time (0 to 90% of

fullscale). Adjusting clockwise decreases response time.

11

Response Ratio Timing

Potentiometer R29

Permits compensation for slight gain changes that may result from adjustment of R20.

At the factory, R29 is adjusted to establish the exact resistance ratio required and is

then secured with locking compound.

12

ALARM 2 Calibration

Adjustment R67

Used for initial calibration of ALARM 2 circuit.

13

ALARM 2 Setpoint

Adjustment R68

Provides continuously variable adjustment of setpoint for ALARM 2 circuit on alarm

accessory, for actuation of external, customer supplied alarm and/or control device(s).

Adjustment range is 0 to 100% of fullscale span.

14

ALARM 2 Deadband

Adjustment R78

Permits adjusting deadband of ALARM 2 circuit from 1% of fullscale (counterclockwise

limit) to 20% of fullscale (clockwise limit). Deadband is essentially symmetrical with

respect to setpoint.

15

ALARM 1 Calibration

Adjustment R63

See item 12 above.

16

ALARM 1 Setpoint

Adjustment R64

See item 13 above.

17

ALARM 1 Deadband

Adjustment R73

See item 14 above.

8

Current Output Zero

Adjustment R1

Use to set zero-level current output, i.e., 4mA for 4 to 20mA board, 0mA for 0 to 20mA

board.

19

Current Output Span

Adjustment R2

Use to set fullscale current output at 20mA for 4 to 20mA or 0 to 20mA board or at

50mA for 10 to 50mA board.

Refer to

Figure 3-2

, page 3-7, for component locations.

Table 3-2.

Model 755A Internal Adjustments