Wall-Mounted Dewpoint Sensor

Product Information Sheet

210-2002



7.

Set Gain to 16, Offset to 0 (the default) and

enter degrees fahrenheit units into the Eng.

Unit field (for example, type DEGF).


8.

To check the sensor input value, press 0 to

go back to the Sensor Control screen and

select 1 -Status.





Set Up Zero Calibration

The Zero Calibration mode allows the user to
calibrate the sensor to a verified zero concentration

of water vapor. For best results, allow the sensor to

warm up for at least 10 minutes.

For zero calibration, all tubing should be connected
between the gas bottle and the sensor inlet flow port.

Before initiating calibration using the sensor keypad,

the gas should be flowed to the sensor at a rate of 80-

100 cc\minute for a period of five minutes.

Observe the following steps to set up Zero Calibration

mode.

1.

Attach the short hose to the bottom port on

the flowmeter.

2.

Attach the long hose to the top port on the

flowmeter.

3.

To ensure the meter is kept in the vertical

position, secure the flowmeter to the side of
the gas bottle using the supplied tie wrap.

4.

Remove the protective cap from the

nitrogen bottle and attach the gas regulator.

5.

Attach the open end of the bottom hose

(located on the flowmeter) to the gas
regulator. Slide the hose far enough on the

gas port to ensure a secure, airtight

connection.

6.

Insert the male luer fitting (located on the

longer hose) into the calibration port,
located on the bottom of the dewpoint

sensor.

7.

Verify that all components are installed

correctly and initiate the calibration process
by turning the knob on the regulator. Turn

the knob until the indicator reaches 7 psi.

8.

Turn the flowmeter knob until the floater

reaches 80-100 cc/minute.

9.

Allow the gas to flow for at least 5 minutes

before proceeding.