Ocx 8800, Instruction manual – Emerson ROSEMOUNT OCX 8800 User Manual
Page 18

Instruction Manual
IM-106-880, Rev 1.0
January 2007
OCX 8800
1-10
NOTE
All static performance characteristics are with operating variables constant. Specifications subject to change
without notice.
Sensors Housing
NEMA 4, IP66 with fitting and pipe on reference exhaust port to
clean, dry atmosphere, two 3/4-14 NPT conduit ports
Electronics Housing
NEMA 4, IP66 with fitting and pipe on reference exhaust port to
clean, dry atmosphere, two 3/4-14 NPT conduit ports
Electrical Noise
EN 61326-1, Class A
Certifications
Sensor Housing
Electronics Housing
Line Voltage
Universal 100 to 240 VAC ±10%, 50 to 60 Hz, no switches or
jumpers required, 3/4-14 NPT conduit port
Pollution Degree
2
Over Voltage Category
II
Relative Humidity
5 to 95% (non-condensing)
Isolated Output
Oxygen
4-20 mAdc, 950 ohm maximum, with HART capability
Combustibles
4-20 mAdc, 950 ohm maximum
Alarm
Alarm output relay - dry contact, form C, 30 mA, 30VDC capacity
Power Consumption
750 W maximum
Specifications
KEMA 04 ATEX 2308
EEx d IIB+H2 T3
0344
II 2 G
2005.1602514
CLASS 1, ZONE 1
Ex d IIB+H2 T3
AEx D IIB+H2 T3
TYPE 4 / IP66
APPROVED
CLASS 1, ZONE 1
AEx D IIB+H2 T3
TYPE 4 / IP66
C
US
2005.1602514
CLASS 1, ZONE 1
Ex d IIB+H2 T6
AEx D IIB+H2 T6
TYPE 4 / IP66
KEMA 04 ATEX 2308
EEx d IIB+H2 T6
IP66
0344
II 2 G
CLASS 1, ZONE 1
AEx D IIB+H2 T3
TYPE 4 / IP66
APPROVED
C
US