beautypg.com

Recharging – Ion Science MVI User Manual

Page 15

background image

MVI MANUAL

Ion Science Ltd


Page 15 of 23

Unrivalled Detection. www.ionscience.com

Battery charging

The MVI incorporates a NiMH battery with a continuous duty cycle of greater than 6 hours’ operation. The
full charge cycle for the battery is 16 hours.

Please note: If the units are charged from a flat battery the unit should not be charged for more than 16
hours continuously.
If the unit requires a top up charging then 2

– 8 hours charging should be sufficient.



Recharging

Switch the instrument off. Connect the MVI battery charger to the instrument. Use only the MVI charger
supplied with the instrument. Use of an alternative charger may damage the instrument and will void the
warranty. The green charging LED lamp located immediately below the ON/OFF switch will light indicating
the batteries are accepting charge.


Only store the unit in a fully charged state.



IMPORTANT NOTE

Only use the Ion Science universal battery charger (Part Number: A-26220) provided with your
instrument.

When received as new or after Service by Ion Science or an approved Ion Science Service Centre,
your Instrument will arrive fully charged.

Do NOT recharge until the “LO BAT” symbol appears. Do

not leave connected to the AC power supply unless charging and certainly no more than 16 hours
from a low battery message.

Excessive charging will dramatically reduce your battery life and will mean the batteries require
replacing frequently.