beautypg.com

Ion Science GasCheck IS User Manual

Page 22

background image

GASCHECK 3000 & 3000IS MANUAL

Ion Science Ltd




Page 22 of 28

Unrivalled Detection. www.ionscience.com

Maintenance













1 Catch

4

Battery compartment

2 Battery compartment cover

5

Serial number

3 Captive screw location (3000is only)

Figure 6 – Replace the batteries (GasCheck 3000 and 3000is shown)



Replacing the batteries















Note: A spare battery holder is supplied with the GasCheck

WARNING

Replace the batteries in the Gas Check 3000is in a safe area where there are

no flammable gases. If you do not, there may be a risk of fire or explosion.

WARNING

Use only dry cell (non-rechargeable AA) batteries or Ion Science AAA Rechargeable Battery

Pack (part No. A-31117) in the GasCheck 3000is. The internal

resistance of unspotted rechargeable batteries is too low to maintain a T4 safety rating.

The recommended batteries are specified in section 2.3 Battery Data.

CAUTION

Ensure that the rubber sleeve on the female part of the battery holder is correctly

fitted when you fit new batteries. If you do not, you could refit the holder

in the GasCheck 3000is with incorrect polarity which will damage the GasCheck 3000is.

Gas Check 3000

1

2

4

5

A

B

3

4

5

Gas Check 3000IS

A

B