Gas selection – Ion Science GasCheck Tesla User Manual
Page 13
GASCHECK TESLA MANUAL
Ion Science Ltd
Page 13 of 25
Unrivalled Detection. www.ionscience.com
Functions
Gas selection
GasCheck Tesla is calibrated using Helium gas at manufacture, however other
gases can be detected and measured using a gas table within the instrument.
Although GasCheck Tesla can't differentiate between gases it can quantify other
gases using response factors stored in its internal gas table. Gas groups 0, 1 2, 3,
and 4 offer a generic sensitivity.
The following list indicates response factors for specific gases however for advice on
gases not listed please contact Ion Science Ltd for advice.
Name
Abbreviation
Minimum sensitivity
cc/sec (ml/sec)
Minimum sensitivity
cc/sec (ml/sec)
Hydrogen
H2
7.7 E-6 cc/sec
0.0000077 cc/sec
Helium
He
1.0 E-5 cc/sec
0.000010 cc/sec
Refrigerant R12
R12
2.7 E-5 cc/sec
0.000027 cc/sec
Refrigerant R1301
R1301
2.4 E-5 cc/sec
0.000024 cc/sec
Refrigerant R134a
R134a
5.8 E-5 cc/sec
0.000058 cc/sec
Refrigerant R22
R22
2.6 E-5 cc/sec
0.000026 cc/sec
Refrigerant R11
R11
3.2 E-5 cc/sec
0.000032 cc/sec
Sulphur Hexaflouride
SF6
2.2 E-5 cc/sec
0.000022 cc/sec
Carbon dioxide
CO2
4.0 E-5 cc/sec
0.000040 cc/sec
Methane
CH4
2.9 E-5 cc/sec
0.000029 cc/sec
Argon
Ar
3.5 E-5 cc/sec
0.000035 cc/sec
Oxygen
O2
2.9 E-4 cc/sec
0.000290 cc/sec
Refrigerant R502
R502
3.0 E-5 cc/sec
0.000030 cc/sec
Refrigerant R404a
R404a
3.2 E-5 cc/sec
0.000032 cc/sec
Refrigerant R407c
R407c
3.3 E-5 cc/sec
0.000033 cc/sec
Refrigerant R410a
R410a
3.2 E-5 cc/sec
0.000032 cc/sec
Refrigerant R507
R507
3.8 E-5 cc/sec
0.000038 cc/sec
To select an alternative gas use the 'Up and Down' keys to select the desired gas
and then press the 'ENTER' key to confirm the selection, at which point a tick will
appear. Press the 'ENTER or Esc' key to exit into the 'Main menu'.
If the gas to be detected does not appear in the instruments internal gas table then
one of the five 'Gas Groups' can be selected. 'Gas Groups' have a variation of
sensitivities. (Group 1 being high sensitivity through to Group 5 being low sensitivity)
Contact Ion Science Ltd for advice. Gas groups allow the instrument to calculate
leak rates with reasonable accuracy.