background image

4

Manual operation

Switch power on (drive rear panel).

Change the set speed by pressing the Ù or Ú key. The 505S speed control ratio is 110:1. This will give a minimum speed of

2rpm for the 220rpm drive and 0.5 rpm for the 55rpm drive.

Change direction by pressing the

CW/CCW key. Check the flashing CW/CCW symbol for actual direction setting. (CW :

clockwise

CCW : counterclockwise).

Select the maximum speed: press the Ù key and the

Max key together. Select the minimum speed: press the Ú key and the

Max key together.

The keypad has a locking facility to avoid resetting or tampering. If the pump is stopped, press

Stop until the padlock symbol

illuminates. If the pump is running, press

Start until the padlock symbol illuminates. All keys will be disabled except for Start

and

Stop. Press these keys until the padlock symbol extinguishes to unlock the keypad.

The pump can be set to automatically restart in its operating state set prior to interruption, or set so that after power is

reconnected the pump will remain stopped. To invoke the Auto-restart facility switch off power to the pump at the mains

supply. Press the

Start key down when the mains supply is switched back on until the ! symbol illuminates. Now press Start

to start the pump. This facility can be cancelled by turning the mains supply off and then pressing the

Stop key whilst turning

the mains supply back on. The

! symbol will not be illuminated.

Press

Start to start the pump. Press Stop to stop the pump.

Error messages

If a fault condition is detected in the drive unit it will stop, all keys will be disabled, and the display will flash:

Er1

Tachometer fault

Er2

Over temperature error

Er3

EEPROM error

Er4

EEPROM read error

Er5

EEPROM write error

Er6

EEPROM exhausted error. There is a maximum number of times the EEPROM can be written to. If Er6 is displayed,

however, the EEPROM must be replaced.

Er9

RAM corruption error

Care and maintenance

The only scheduled maintenance of the 505S is to inspect the motor brushes and to replace them before their length is less

than 6mm (1/4"). The life of the brushes will depend on the duty of the pump, which is expected to be at least 10,000 hours at

maximum speed. When the pump needs cleaning, remove the pumphead and use a mild solution of detergent in water. Do not

use strong solvents.
If the gearbox is rebuilt you should use 15 ml of the recommended lubricant, which is RD105, this is a SAE 30 mineral oil loaded

with molybdenum disulphide to form a soft fluid grease.

Specification

Maximum rotor speed

55rpm, 220rpm

Voltage/frequency

100-120/220-240V 50/60Hz

Control range

110:1

Power consumption

100VA

Shaft Torque

2.2Nm

Operating temperature range

5 to 40C

Storage temperature range

-40C to 70C

Weight

7.7kg (17Ib)

Noise

<70dBA at 1m

Standards

IEC 335-1, EN60529 (IP31)

Machinery Directive 98/37/EC EN60204- 1

Low Voltage Directive 73/23/EEC EN61010- 1

EMC Directive:89/336/EEC EN50081-1 / EN50082-1

501RL Pumphead

The 501RL pumphead has two spring-loaded working rollers, which automatically compensate for minor variations in tubing

wall thickness, giving extended tube life. The 501RL is set during manufacture to accept tubing with wall thicknesses of between

1.6mm and 2.0mm, and internal diameters of up to 8.0mm. It is equipped with a "tool lockable" guard for increased safety. This

should be locked shut whilst the pump is in use.