beautypg.com

4 operating parameters – Watson-Marlow 700 series User Manual

Page 3

background image

ATEX Directive: Watson-Marlow 700 series pumpheads User Manual

3

4 Operating parameters

The following tube materials can be used with 701RA pumpheads. Bore sizes range
from 9.6mm to 25.4mm, with a wall thickness of 4.8mm.

Tubing: working temperature range
Marprene

5C to 80C

Bioprene

5C to 80C

Pumpsil platinum-cured silicone -20C to 80C
STA-PURE PCS

0C to 80C

STA-PURE PFL

0C to 80C

Neoprene

0C to 80C

PVC

0C to 80C

The following parameters define the boundary of the safe working envelope - these
values must not be exceeded (ATEX compliance will be invalidated).

When two or more items of ATEX equipment are combined, the permissible operat-
ing envelope will be determined by the narrowest range after considering all values
for a given parameter.

701RA pumpheads
Ambient temperature range

5C to 40C

Max peak pressure (701RA, 701RXA,

7014REA and 701REXA)

2 bar

Max peak pressure (701RGA)

4 bar

Max continuous speed

360 rpm

Corrosion resistance

See 10 Materials

Tube life

See 7 Tube life

WARNING! Do not run dry for excessive periods. Roller and tubing
temperatures can exceed normal operating range.

WARNING! Do not run pumphead against a closed or restricted dis-
charge pipe. This can lead to excessive roller and tubing temperatures
and pressures in excess of the limits in the table above. Excessive
pressure may cause tube failure.

This manual is related to the following products: