8 114dv performance data – Watson-Marlow 120F/DV User Manual

Page 25

background image

Watson-Marlow 114DV pumphead: operating instructions

7

3.8 114DV
Performance data

Flow rates are normalised test values obtained
using new tubing with the pumphead rotating clock-
wise pumping water at 20C with negligible inlet and
discharge pressure. Actual flow rate achieved may
vary because of changes in temperature, viscosity,
inlet and discharge pressures, system configuration
and tubing performance against time. Flow rates
may also vary due to normal manufacturing
tolerances of the tubing. These tolerances will make
flow rate variance more pronounced at smaller bore
sizes.

For precise and repeatable performance it is
important to determine flow rates under
operating conditions for each new piece of
tubing.

Note: Flow rates quoted have been rounded for
simplicity, but are accurate to within 5%, well
within the normal tubing tolerance variation of flow
rate. They should therefore be taken as a guide.
Real flow rates in any application must be
determined empirically.

120F: All tubing materials (ml/min)

Fixed

speed

0.5

mm

0.8

mm

1.6

mm

2.4

mm

3.2

mm

4.0

mm

4.8

mm

10 rpm

0.2

0.4

1.4

2.9

4.7

6.7

8.5

17 rpm

0.3

0.7

2.4

4.9

8

11

14

31 rpm

0.6

1.2

4.3

9.0

15

21

26

52 rpm

1.0

2.1

7.3

15

25

35

44

220 rpm

4.4

8.8

31

64

100

150

190

120S: All tubing materials (ml/min)

Speed

0.5mm

0.8mm

1.6mm

2.4mm

3.2mm

4.0mm

4.8mm

1-200

rpm

0.02-4

0.04-8

0.14-28

0.29-58

0.47-94

0.67-130

0.85-170

120U: All tubing materials (ml/min)

Speed

0.5mm

0.8mm

1.6mm

2.4mm

3.2mm

4.0mm

4.8mm

0.1-200

rpm

0.002-4

0.004-8

0.01-28

0.03-58

0.05-94

0.07-130

0.09-170

Pressure and flow performance curve

Marprene tubing, 1.6mm wall, 100 rpm, CCW
rotation. Note: The 120S and the 120U must be run
counter-clockwise to achieve pressures up to 2 bar.
The 120F is a transfer-only, 0 bar pump.

suction (bar) pressure (bar)

0

1

2

Flow rate: ml/min

0

20

40

70

80

100

10

30

50

60

90

-0.5

-0.3

0

-0.1

-0.2

-0.4

0.8 x 1.6

1.6 x 1.6

4.8 x 1.6

4.0 x 1.6

3.2 x 1.6

2.4 x 1.6

114DV, standard occlusion