Watson-Marlow Alitea 114 pumpheads: operating instructions

9

7 Tube holder positioning

The pumphead can be adjusted to accommodate 1.6mm wall tubing in sizes from
0.5mm bore to 4.8mm bore.

Inner position, for small tubing

Outer position, for large tubing

With the smaller bore tubes of 0.5mm, 0.8mm and 1.6mm the inner position must
be used to prevent the risk of tube slipping through the clamps and wandering
across the rollers causing premature tube rupture.

With the larger bore tubes of 4.0mm and 4.8mm the outer position must be used to
prevent the flow rate being excessively reduced.

For tubing bores of 2.4mm and 3.2mm either setting may be used, as appropriate
for the application. The inner setting will clamp the tube harder, reducing tube slip
but has the potential to marginally reduce flow rate. The outer setting will optimise
flow rate but the risk of tube slip is increased.

114DV, 114DVP

Tube holder position

Tube bore size

0.5mm

0.8mm

1.6mm

2.4mm

3.2mm

4.0mm

4.8mm

Inner

Outer