beautypg.com

Part 2: operation overview, Flow rates – Watson-Marlow 603S/R User Manual

Page 6

background image

Page 6

Part 2: Operation

Overview

Start the pump by turning the forward/off/reverse switch on the left hand side

of the front panel to the required direction of rotation. The preferred direction

of rotation is clockwise (with fluid entering at the bottom right of the

pumphead and exiting at the top right of the pumphead), which will ensure the

longest possible tube life. To operate against higher pressures, use

anticlockwise rotation.
The pump can be primed rapidly at its maximum speed by depressing the

max switch on the front panel. When released the switch will return to its

normal position.
The speed control of the 603S/R is 20:1 giving a speed range from

approximately 8rpm to its rated maximum speed of 165rpm. The digital read-

out shows true motor speed in revolutions per minute. A locking knob is fitted

to the potentiometer which should be pulled out to unlock, turned to give the

appropriate speed setting, and pushed in to relock.
To stop the pump, turn the forward/off/reverse switch to its central off

position. To change the direction of flow, turn the forward/off/reverse switch to

its central off position until the pumphead rotor stops, and then turn it to the

other direction of rotation.

Flow rates

The flow rates given below were obtained using Watson-Marlow 3.2mm wall

thickness Marprene tubing pumping water at 20C with negligible suction and

delivery pressures (unless otherwise stated). Where an application is critical,

the flow rate should be determined under operating conditions. The important

factors are suction and delivery pressures, temperature and fluid viscosity.

603 S/R flow rates

Minimum flows 5% percent of rates given

Tube #

123

26

73

82

184

Tube Bore

4.8mm

6.4mm

9.6mm

12.7mm

15.9mm

3/16"

1/4"

3/8"

1/2"

5/8"

Litre/min

1.5

2.3

4.2 7.4

11

US gpm

0.4

0.6

1.1 2.0

2.9


603S/R flow rates against suction lift and delivery pressure. Suction lift figures

obtained with zero delivery pressure. Delivery pressure figures obtained with

zero suction lift.

Suction (feet of water)

Pressure (bar)

Suction (mmHg)

Pressure (psig)

15.9

12.7

6.4

0

2

4

6

8

10

12

lit

re

/m

in

15

45

12

16

8

0

30

4

300

400

200

0

1

2

3

100

9.6

Clockwise