Part 1: 701S/R Drive

Siting

The 701S/R can be operated at ambient temperatures from 5C to 40C.

Storage temperatures from -40C to 70C are permissible, but allow time for

acclimatisation before use. The pump should be positioned to allow a free

flow of air around it.
Should the pump fail to operate, check that mains electricity is available at the

unit, that the fuses are intact and that the pump is not stalled by incorrect

fitting of tubing.
A drain hole (tapped 1/4" BSP) is provided at the left hand side of the

pumphead base which is plugged during assembly. If required, the plug can

be removed and a drain hose connected.

Installation

The 701S/R is suitable for single phase mains electricity only. Ensure that the

supply voltage and frequency correspond with those marked with a white dot

on the rear panel, (220V 50/60Hz, 120V 50/60Hz or 100V 50/60Hz).

Supplies

Remove the small plate on the rear panel to gain access to the terminal block.

Route the mains cable through the entry point to the right of the recess and

couple the cable to the terminal block in accordance with the instructions

printed on the rear panel. Two alternative connectors are supplied to screw

into the entry point gland. One accepts 20mm rigid or flexible conduit, and the

other accepts a three core 0.75 square mm PVC sheathed mains cable (via

the screwed adaptor provided) so that a flying lead can be used. Securely

replace the transparent plate and gasket over the recess.

Back panel

Power cable
5 to 8mm outside diameter

M20 for direct conduit connection


Ingress protection standard will be compromised if transparent plate is

not properly replaced.

There are dangerous voltages (at mains potential) inside the motor. If

access is required, isolate the pump from the mains before removing

the cover.