Specification

Nominal maximum rotor speed

360rpm

Speed control ratio

50:1

Maximum power consumption

750VA

Operating temperature range

5C to 40C

Storage temperature range

-40C to 70C

Weight 26kg

Noise ( one pumphead )

<78dBA at 1m

Standards

IEC 335-1, EN60529 (IP55)

Machinery Directive: 89/392EEC

and EN602041

Low Voltage Directive: 73/23/EEC

and EN61010-1

EMC Directive: 89/336/EEC

and EN50081-1/EN50082-1

Care and maintenance

The 701S/R uses a permanent magnet direct current motor with a pulse-

width-modulated speed controller which provides black commutation. The

only scheduled maintenance required for the 701S/R is inspection of the

motor brushes and their replacement before their length is less than 10mm.

The life of the brushes will depend on the duty of the pump, but is expected to

be at least 2000 hours at maximum speed.
The sun gear of the gearbox in the 701R and 701RX pumpheads should be

lightly greased with a quality gear grease every one thousand hours. If

harmful liquids are spilled on to the pump, the case and pumphead should be

thoroughly cleaned with detergent and water. Strong solvents should not be

used. The sun gear of the gearbox in the pumphead should be lightly greased

with a quality gear grease after the pumphead has been cleaned.