Rotor removal

Remove any tubing from the pumphead.

Open the flexible rotor cap in the centre of the rotor.

Undo and withdraw the central locating screw using a slotted screwdriver.

Pull the rotor hub off its dogged shaft.

Between the hub and the shaft is a split collet. If the collet
is retained by the shaft, pull it off, loosening it if necessary
by tapping it lightly. Avoid levering it off using a
screwdriver or other tool. If the collet is retained within
the hub, remove it, loosening it if necessary by reinserting
the central locating screw a turn or two and tapping the
screw head lightly.

Rotor replacement

Re-locate the split collet onto the drive shaft, rotating it until it fully engages the dog. Fit
the rotor body over the drive shaft as one unit.

Open the flexible rotor cap in the centre of the rotor. Use a slotted screwdriver to tighten
the central locating screw to a torque of 3Nm (2.2 lb-ft) to prevent collet slip during
operation. When fitted correctly, the tube guide rollers should align with the outer face of
the track. Close the flexible rotor cap.

Close the guard and ensure that the rotor is clear of the guard by observing the first few
rotor rotations.