Example 2 - Excess of volume
If we have a flowmeter body which specifies i/l = 1.985, and when we check the volume
of a batch, we find that instead of having 100 litres as programmed, we only have 105
litres (5% more) and we must apply the following correction:

i/l = Original Pulses per litre Factor

= 1.985

V = Expected Volume

= 100

Vr = Real Volume

= 105

i/l n = New Pulses per litre Factor

= ? (1.887)

6 HART

COMMUNICATION

The CH420R transmitters have a HART communication MODEM
The details of the characteristics with respect to the HART communication are available in
the appropriate Field Device Specification

Resume of the principal communications characteristics:

Manufacturer, Model and Revision

Tecfluid S.A., CH420R, Rev. 0

Device type

Transmitter

Hart Revision

6.0

Device Description available

No

Number and type of sensors

1, external

Number and type of actuators

0

Number and type of host side signals

1, 4 – 20 mA analog

Number of Device Variables

2

Number of Dynamic Variables

1

Mappable Dynamic Variables

No

Number of Common Practice Commands

13

Number of Device Specific Commands

2

Bits of Additional Device Status

12

Alternative operating modes?

No

Burst mode?

No

Write Protection?

Yes

i/l n = i/l

V

Vr