8

Example:

Suppose a pipe of 300 mm inner diameter (DN300) and 5.5 mm thick. In the table we can
see that the distance that the insertion fitting must extend above the exterior wall is H =
60,5 - s = 60,5 -5,5 = 55 mm.

The values in the table are calculated with the gasket supplied with the instrument, which
is 3 mm thick. If the thickness of the gasket is changed, the value of H will change.

The equation to calculate H’ for a gasket of thickness d is the following:

H’ = H + 3 - d

In the previous example, if the gasket was 5 mm thick, the distance that the fitting should
extend above the exterior wall would be H’ = 55 + 3 - 5 = 53 mm.

DN

Sensor

length

Fitting
length

C

(H)

80

101

93

10

88-s

100

12,5

85,5-s

125

15,5

82,5-s

150

19

79-s

200

25

73-s

250

31

67-s

300

37,5

60,5-s

350

44

54-s

400

50

48-s

500

206

145

62,5

140,5-s

600

75

128-s

700

87,5

115,5-s

800

100

103-s

900

112,5

90,5-s

1000

125

78-s

1200

356

190

150

203-s

1400

175

178-s

1600

200

153-s

1800

225

128-s

2000

250

103-s

Qnom

m

3

/h

90,5

141,3

220,9

318,1

565,5

883,6

1.272.,3

1.731,8

2.261,9

3.534,3

5.089,4

6.927,2

9.047,8

11.451,1

14.137,1

20.357,5

27.708,8

36.191,1

45.804,4

56.548,7

Threaded fitting

Flange fitting