beautypg.com

4-wire connection, Technical characteristics, 1 power supply – Tecfluid TH6 For AD Series User Manual

Page 7: 2 outputs, 3 general characteristics

background image

7

7 TECHNICAL

CHARACTERISTICS

7.1 Power

supply

2 wire.

Minimum voltage (TH6):

0.02 Z + 12 (Volt) (Z is the load in the current
loop in Ohm).

The minimum value is 12 VDC for Z=0 Ohm.

Minimum voltage (TH6H):

0..02 (Z+Rext) + 14 (Volt) (Z is the load in the
current loop in Ohm).

The minimum value is 18 VDC for Z=0 Ohm and
Rext=200 Ohm.

Maximum voltage:

36 VDC (for the Ex version see point 8.3)

Consumption:

maximum

20

mA

7.2 Outputs

Analog output:

4 - 20 mA, factory calibrated

Maximum load in the 4-20 loop:

1.1 k

 (at 36 VDC supply voltage)

7.3 General

characteristics

Housing ingress protection:

IP65

Cable glands:

M12 x 1.5

Ambient temperature:

-5 ...+70 ºC

Precision (analog output respect
the magnetic field):

< 0.6 %

6 4-WIRE

CONNECTION

If Direct Current power supply for the transmitter is not available in the installation, it will
be necessary to incorporate an additional power supply as in the following figure.

Power

supply

Mains

This manual is related to the following products: