Once the password is written, press "Enter" or "Open", and the Calibration or Programming
tab will open. At the bottom of each section the text "Calibration tab open" or "Programming
tab open" will be displayed
To enter the Installation window, just click the corresponding tab.

In the calibration window a complete re-calibration of the transmitter according to the scale
plate can be done.
The first step is to choose the scale units, then the specific units.
In the combo N. points the number of points with which the calibration will be performed is
selected. The minimum is 10 and the maximum 16.
With these data, the boxes P0 ... P10 to P16 are filled with the values of the scale in which
the adjustment will be made.
To perform the calibration, the instrument must be in its operating position.
Depending on the instrument in question, move the float or the disc until the needle points
each calibration point, and press the "Calibrate" button of the corresponding point.
Once calibrated every point, to send all the data to the transmitter TH7, click the "Send"
button. The data is then stored in the transmitter memory.