5

OPERATION

Wait 20 minutes after the test subject smokes, eats or drinks

before testing. Failure to wait 20 minutes can give inaccurate BAC

readings and damage the sensor.
Step 1)
Place a fitted cover over the mouthpiece each time you test

another individual.

Step 2) If it has been more than 24 hours since the device was last

used,

prime the sensor by blowing into the mouthpiece for 4

to 5 seconds

before turning on the device.

Step 3) PRESS the black Power On/Off button located on the front

of the unit. A single beep will sound and the device will count

down from 60 as it prepares the sensor for testing.

Note: The device will restart a count down from 10 if the sensor is not

ready within the first 60 second countdown.

Step 4) WAIT until the RED LIGHT IS SOLID RED and you hear a

SINGLE BEEP.

Note: the device will not shut off until a test is performed OR until 15

seconds of non use elapses.

Step 5) Take a DEEP BREATH AND BLOW STEADILY (for at least

5 seconds) until a

BEEP sounds to signal completion of the

testing.