Specifications, Verification and calibration, Calibration of the alcohawk – Q3 Innovations AlcoHAWK SLIM User Manual
Page 10
8
SPECIFICATIONS
Dimensions
5” x 1.75” x .75” (12.7 cm x 4.45 cm x 1.91 cm)
Weight
1.8 oz (51 g) without batteries; 3.4 oz (96 g) with batteries
Housing
Shock resistant, molded plastic
Power Supply
Two (2) AA alkaline batteries
Battery Life
150 tests
1
Sensor Technology
Sensitive Semiconductor Sensor
Blowing Time
5 seconds
Response Time
5 seconds
1
Digital Display
2 digits (.00% BAC)
Operating Temperature
10 - 40°C (50 - 104°F)
Detection Range
.00% - .40% BAC
Air Sample
5 seconds
Operation
1 button
Warranty
1 Year Limited Warranty
1
Estimates, results may vary.
VERIFICATION AND CALIBRATION
Calibration of the AlcoHAWK
®
Slim occurs during manufacturing using
advanced alcohol simulation equipment. The manufacturer passing
known alcohol concentrations through the sensor to set baseline values
for testing. During testing, the unit compares users’ breath samples
to these baseline values. If the verification determines that the BAC is
not consistent with actual BAC levels, a technician calibrates the unit.
The accuracy of breath alcohol testers can fluctuate after six to twelve
months, depending on frequency of use and operating conditions.