Step 6) For more testing, wait one (1) minute and repeat steps 1 – 5.

READ THE RESULTS

The result is an estimate of the subject’s BAC. A breath alcohol tester

cannot determine the exact BAC in the blood of the subject. Correlation

between BrAC and BAC depends on many factors, such as temperature

and health conditions. Only use this device to screen for the presence

of alcohol.
Many people consider a 0.02% or greater a positive test for alcohol in

“zero tolerance” screening. Most states have 0.08% as the legal BAC

limit for tasks such as operating a motor vehicle. Some states prohibit

driving commercial vehicles or any vessel with a 0.04% BAC or higher.

Many states prosecute individuals for driving under the influence at any

level.

ERROR MESSAGES

Error Message

Cause and Course Of Action

SNR

The sensor is not ready for testing. This can occur

if the unit has not been used for several days and if

the unit has not been primed. If the SNR message

is displayed, press the Power On/Off button

once to turn off the unit and follow the operating

instructions again.

Sht

The user had a short exhale time during the test,

which likely means the user did not exhale steadily

and continuously for at least 5 seconds. Repeat

test.

6