Step 6) For more testing, wait one (1) minute and repeat steps 1 – 5.
READ THE RESULTS
The result is an estimate of the subject’s BAC. A breath alcohol tester
cannot determine the exact BAC in the blood of the subject. Correlation
between BrAC and BAC depends on many factors, such as temperature
and health conditions. Only use this device to screen for the presence
of alcohol.
Many people consider a 0.02% or greater a positive test for alcohol in
“zero tolerance” screening. Most states have 0.08% as the legal BAC
limit for tasks such as operating a motor vehicle. Some states prohibit
driving commercial vehicles or any vessel with a 0.04% BAC or higher.
Many states prosecute individuals for driving under the influence at any
level.
ERROR MESSAGES
Error Message
Cause and Course Of Action
SNR
The sensor is not ready for testing. This can occur
if the unit has not been used for several days and if
the unit has not been primed. If the SNR message
is displayed, press the Power On/Off button
once to turn off the unit and follow the operating
instructions again.
Sht
The user had a short exhale time during the test,
which likely means the user did not exhale steadily
and continuously for at least 5 seconds. Repeat
test.