Inova Labs XYC100B-P4L User Manual
Page 30

28
The LifeChoice Activox Portable Oxygen Concentrator Model XYC100B-
P4L
is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the Model
XYC100B-
P4L
can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment
(transmitters) and the Model XYC100B-
P4L
as recommended below, according to
the maximum output power of the communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended
separation distance ‘d’ in meters can be estimated using the equation applicable
to the frequency of the transmitter, where ‘P’ is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80MHz and 800MHz, the separation distance for the higher frequency range
applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects, and people.
Recommended Separation Distances Between Portable and Mobile
RF Communications Equipment and The LifeChoice
®
Activox Portable
Oxygen Concentrator Model XYC100B-P4L
The LifeChoice
®
Activox Portable Oxygen Concentrator Model XYC100B-
P4L
is intended for use in the electromagnetic environment specified below. The
customer or the user of the Model XYC100B-
P4L
should assure that it is used
in such an environment.
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
Separation Distance According to the Frequency of Transmitter (m)
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.2
1.2
2.3
10
3.7
3.7
7.4
100
12
12
23
Rated Maximum
Output Power of
Transmitter (W)
150kHz to 80MHz
80MHz to 800MHz 800MHz to 2.5GHz
d=1.17 √P
d=1.17 √P
d=2.33 √P
Emissions Test
Compliance
Electromagnetic Environment - Guidance
The Model XYC100B-P4L uses RF energy only for
its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference
to nearby electronic equipment.
RF Emissions, CISPR 11
RF Emissions, CISPR 11
Harmonic Emissions,
IEC 61000-3-2
Voltage Fluctuations/Flicker
Emissions, IEC 61000-3-3
Group 1
Class B
Class A
Complies
The Model XYC100B-P4L is suitable for use
in all establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.