Inova Labs XYC100B-P4L User Manual
Page 28

26
The LifeChoice Activox Portable Oxygen Concentrator Model XYC100B-P4L
is intended for use in the electromagnetic environment specified below. The
customer or the user of the Model XYC100B-P4L should assure that it is used in
such an environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
Immunity Test
IEC 60601
Compliance
Electromagnetic
Test Level
Level
Environment - Guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
+/-6 kV contact
+/-8 kV air
+/-6 kV contact
+/-8 kV air
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Power frequency magnetic
fields should be at levels
characteristic of a typical
commercial or hospital
environment.
3 A/m
3 A/m
Power Frequency
(50/60Hz)
Magnetic Field
IEC 61000-4-8
Main power quality
should be that of a typical
commercial or hospital
environment.
+/-2 kV
for power
supply lines
+/-1kV for input/
output lines
+/-2 kV
for power
supply lines
Not applicable
Electrical Fast
Transient/Burst
IEC 61000-4-4
Main power quality
should be that of a
typical commercial
or hospital
environment.
+/-1 kV
line-to-line
+/-2 kV
line-to-earth
+/-1 kV
line-to-line
+/-2 kV
line-to-earth
Surge
IEC 61000-4-5
Main power quality should be
that of a typical
commercial or hospital
environment. If the user of
the Model XYC100B-
P4L requires continued
operation during power
mains interruptions, it is
recommended that
the Model XYC100B-
P4L be powered from an
uninterruptible power supply
or a battery.
< 5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in U
T
)
for 5 seconds
< 5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in U
T
)
for 5 seconds
Voltage dips,
short interruptions
and voltage variations
on power supply
input lines
IEC 61000-4-11
Medical electrical equipment needs special precautions regarding EMC and needs
to be installed and put into service according to the EMC information provided in
the accompanying tables.
EMC INFORMATION