Operating Instructions (Contd)

Loading the Pump

A hinged black plastic compression clamp holds the tubing against the

rollers. Release the clamp by pushing it and rotating it away from the

rollers. Insert one or two tube segments and close the clamp tightly until it

snaps into position. See page 6 for more detailed instructions and diagrams.

Setting the Direction

A toggle switch has been provided for clockwise or counter clockwise

rotation.

Adjusting the Flow Rate

Use the speed range switch x1, x2, and the main speed knob to adjust

flow rate.

F

Fllo

ow

w R

Ra

atte

e p

pe

err ttu

ub

be

e,, m

mll//m

miin

n..

(Below values are for a single tube. Using two tubes in parallel will double

the flow rate.Values are approximate.)

Use a stop watch and measured container to precisely determine the

actual flow rate.

MPII Flow Rates in ml/min

Switch

With One Tube

With Two Tubes

Setting

Min.

Max.

Min.

Max.

x1

0.8 ml/min

7.00 ml/min

1.6 ml/min

14.00 ml/min

x2

1.5 ml/min

12.25 ml/min

3.0 ml/min

24.50 ml/min