beautypg.com

Operating instructions – Hoefer PP24 User Manual

Page 6

background image

4

H

a

r

v

a

r

d

A

p

p

a

r

a

t

u

s

M

P

I

I

M

in

i-

P

e

r

is

t

a

lt

ic

P

u

m

p

Publication 5392-001-REV-A

Operating Instructions

Description

The Harvard MPII Mini-Peristaltic pump takes only one size of tubing, 1.6

mm ID x 3.2 mm OD (1/16 x 1/8 in). It can be used with either a single tube

or two tubes simultaneously.

Two front panel controls provide flow rates from approximately 0.8 to

12.25 ml/min. The control knob provides variable adjustment from 0 to

100% of the selected flow rate range.The second control is a two position

toggle switch marked x1, x2 which selects low or high flow rates, see table

to right.

The easy-loading four-roller pump head is on top of the stout metal box.The

back of the pump head effortlessly rotates into an ‘open’ position and either

one or two tubes can be dropped into slots. The loaded section simply

rotates back against spring loaded jaws and locks into place. The tubing is

automatically in proper wiping contact with the pump head rollers. Each

Pump is provided with a 12.5 mm (0.5 in) rod clamp on the back so that

multiple pumps can be mounted vertically on a lattice rod.

Location Requirements for the MP II Mini-Peristaltic Pump

• A Sturdy, level, clean and dry surface

• Minimum of one inch (2 cm) clearance around the pump

• Adequate power supply

• Room temperature 4°C to 40°C (40°F to 104°F)

• Relative humidity 20% to 80%

• A well ventilated room

Power Requirements

The MP II Mini Peristaltic Pump requires 12 VDC at 800 mA. The power

receptacle has a 2.5 mm center pin, which is the positive polarity. One (1)

universal power supply is provided for 120/230 VAC operation. Plug the out-

put connector into the rear receptacle of the Mini Peristaltic pump.The uni-

versal power transformer automatically adjusts the supply voltage. The

North American plug can be replaced with a similar power cord with an IEC

320 plug. Observe the color coding of the international line cord.

Brown – High

Blue

– Neutral

Green

– Ground (earth)

Tubing

The pump is designed for silicone rubber tubing 1/16 inch I.D. x 1/8 inch

O.D. only.The pump is supplied with two short lengths of tubing with end

connectors fitted. If other tubing materials are used, please verify pump

flow rates. Excessively hard tubing may overload the pump motor and

cause if to ‘stall’.