7. MAINTENANCE

Use 70% ethanol and a soft cloth or lint free tissue to wipe the external pipette surfaces
clean.

To clean in the interior of the pipette:

1. Hold down the tip ejector.
2. Place the pronged end of the wrench between the tip ejector and the tip ejector

collar to release the locking mechanism.

3. Carefully release the tip ejector and remove the collar.
4. Using the wrench, turn the tip cone counter clockwise to remove the cone.
5. Wipe the piston, O-ring and the tip cone with 70% ethanol. Note: Pipettes with

volumes of 10 µl or less have a fixed O-ring and cannot be accessed.

6. Grease the piston with a light coating of silicone.
7. Reassemble the pipette and then depress the thumb button several times to

lubricate the piston to make sure it does not bind.

8. Check calibration.

8. TROUBLESHOOTING GUIDE

Problem Possible

Cause Solution

Tip incompatibility

Try a different brand of tip

Liquid left inside the tip

Plastic tip not uniform

Attach a new tip

Tip incorrectly attached

Reattach tip

Tip cone fouled

Clean tip cone

Pipette has material inside
cone

Clean and grease O-ring
and piston; clean tip cone

O-ring not positioned
correctly or damaged

Replace O-ring

Leakage or pipette volume
too small

Calibration incorrect

Recalibrate pipette

Piston damage

Clean and grease O-ring
and piston; clean tip cone

Solvent contamination

Clean and grease O-ring
and piston; clean tip cone

Push button jammed or
off center

Calibration incorrect

Recalibrate pipette

Pipette not aspirating

Material blocking cone tip

Clean and grease O-ring
and piston; clean tip cone

Tip ejector jammed

Ejector collar has debris

Clean tip cone and ejector
collar