9. Compare the calculated Accuracy % and the Precision % values to the

specifications table below. If the result is not within the specifications
the pipette should be recalibrated.

Pipette Volume Range Volume

Accuracy ±

Precision ±

0.1-2.5µl 2.5µl

2.50%

2.00%

1.25µl

3.00%

3.00%

0.25µl

12.0%

6.00%

0.5-10µ 10µl

1.00%

0.80%

5µl

1.50%

1.50%

1µl

2.50%

1.50%

2-20µl 20µl

0.90%

0.40%

10µl

1.20%

1.00%

2µl

3.00%

2.00%

5-50µl 50µl

0.60%

0.30%

25µl

0.90%

0.60%

5µl

2.00%

2.00%

10-100µl 100µl

0.80%

0.15%

50µl

1.00%

0.40%

10µl

3.00%

1.50%

20-200µl 200µl

0.60%

0.15%

100µl

0.80%

0.30%

20µl

3.00%

1.00%

50-200µl 200µl

0.60%

0.15%

100µl

0.80%

0.30%

50µl

1.00%

0.40%

100-1000µl 1000µl

0.60%

0.20%

500µl

0.70%

0.25%

100µl

2.00%

0.70%

200-1000µl 1000µl

0.60%

0.20%

500µl

0.70%

0.25%

200µl

0.90%

0.30%

1-5 ml

5 ml

0.50%

0.15%

2.5 ml

0.60%

0.30%

1 ml

0.70%

0.30%

6.2 Recalibration Procedure

1. Place the calibration tool on the calibration adjustment lock under the

thumb button of the pipette.

2. Turn the adjustment lock counterclockwise to decrease volume and

clockwise to increase volume.

3. Recheck the calibration using the above procedure.
4. Readjust the pipette until the calibration is correct.