Hoefer HE-PLUS System User Manual
Page 12
![background image](https://www.manualsdir.com/files/789000/content/doc012.png)
•
p2
1
Make 500 ml of either 1X TAE or 1X TBE
electrophoresis buffer.
2
Weigh an appropriate quantity of agarose (see
Table 1) and place it into a 250 ml flask. Add a
sufficient quantity of either 1X TAE or 1X TBE buffer
(prepared in step 1) to achieve a final volume of
100 ml agarose solution.
Table 1: Gel Concentrations and Resolving Ranges
Concentration
Agarose (g)
Efficient Range
of Agarose in
per 100 ml
of Separation
Gel (%w/V)
Buffer
of Linear DNA (Kb)
0.3
0.3
5 – 60
0.6
0.6
1 – 20
0.7
0.7
0.8 – 10
0.9
0.9
0.5 – 7
1.2
1.2
0.4 – 6
1.5
1.5
0.2 – 3
2.0
2.0
0.1 – 2
Table taken from Sambrook, J., Fritsch, E.F., & Maniatis, T. (1989)
Molecular Cloning, A Laboratory Manual, 1, 6.8 613.
See also other documents in the category Hoefer Equipment:
- IEF100 (65 pages)
- IEF100 (65 pages)
- SE900 (37 pages)
- SE260 (31 pages)
- SE260 (31 pages)
- SE300 miniVE (37 pages)
- SE300 miniVE (37 pages)
- SE300 miniVE (37 pages)
- SE410 (51 pages)
- SE410 (50 pages)
- SE640 (51 pages)
- SE640 (50 pages)
- SE275 (21 pages)
- SE245 (13 pages)
- SE675 (27 pages)
- SG100 (9 pages)
- SG500 (12 pages)
- SQ33 Sequencer (30 pages)
- SQ33 Sequencer (30 pages)
- SQ33 Sequencer (30 pages)
- HE33 (26 pages)
- HE33 (26 pages)
- HE99X (26 pages)
- HE99X (26 pages)
- HE-PLUS System (26 pages)
- SUB Series (31 pages)
- SUB Series (31 pages)
- TE70X (31 pages)
- TE70X (31 pages)
- PR648 (11 pages)
- PR150 (9 pages)
- PR645 Immunoblot XL (21 pages)
- TE22 (35 pages)
- TE22 (35 pages)
- TE62 (35 pages)
- TE22 (31 pages)
- DQ300 (30 pages)
- DQ300 (30 pages)
- UVIS20 (21 pages)
- UVIS20 (21 pages)
- SP-2001 Vision Life Spectrophotometer (66 pages)
- Single Channel Variable Pipette (8 pages)
- PR250 (18 pages)
- PP24 (10 pages)