Ground resistance, Leakage current, Isolation test – Fluke Biomedical ESA180 User Manual
Page 21: Current capacity, Current consumption, Environmental

Introduction and Specifications
Specifications
1
1-11
Ground Resistance
Range .................................................................... 0 – 2.00
Ω
Accuracy ................................................................
±
1 % of reading,
±
1 LSD
Range .................................................................... 2.01 – 19.99
Ω
Accuracy ................................................................
±
3 % of reading,
±
1 LSD
Current source ....................................................... 10 mA dc
Type....................................................................... 4-wire bridge
Leakage Current
Range .................................................................... 0 – 1999
μ
A
Accuracy:
DC and 25 Hz to 1 kHz ......................................
±
1 % of reading,
±
3 LSD
1.0 kHz to 100 kHz ............................................
±
2.5 % of reading,
±
3 LSD
100 kHz to 1 MHz ..............................................
±
5 % of reading,
±
3 LSD
Type measurement............................................ True rms
Impedance ......................................................... 1000
Ω
, AAMI ESI 1993
Isolation Test
Voltage................................................................... 110 % Line voltage
±
5%
Current................................................................... limited by 120 k
Ω
resistor
Current Capacity
Line 90 to 140 V ac................................................ 20 A max. on time 5 min.
15 A max. on time 30 min.
Line over 140 V ac................................................. 10 A
Current Consumption
ESA 180................................................................. 0.1 A @ 120 V – 60 Hz
Environmental
Operating Temperature ......................................... 15 – 40
°
C
Storage Temperature............................................. -20 – 65
°
C
Relative Humidity................................................... 90 % max.
Mains voltage range .............................................. 90 to 240 V ac